China Cheap price Electrode Wire - 3M Pre-Wire Disposable Neutral Electrode – Quanding

Short Description:


Product Detail

Product Tags

Related Video

Feedback (2)

By using a full scientific excellent administration method, great quality and fantastic religion, we get good reputation and occupied this discipline for Aed Pads On Infant , Defib Pad Placement , Cvs Tens Unit , Should you be interested in almost any goods, remember to truly feel totally free to get in touch with us for further facts or be sure to deliver us email right, we will reply you in just 24 hrs as well as the finest quotation are going to be provided.
China Cheap price Electrode Wire - 3M Pre-Wire Disposable Neutral Electrode – Quanding Detail:

Product Specification 

For Adult For Child Infant
QD202B-A QD202B-P QD202B-1
Bipolar /Split Monopolar/single Bipolar/Monopolar
102*202mm 90*148mm 53.5*124mm
Packaging 1pc/foil bag350pcs/carton  
Carton Size 45*45*28CM GW: 15.4kg  
Certification ISO13485,CE  

Storage

 

The product must be stored in its original package at the environmental conditions (temperature and relative humidity) specified on the pouch’s label. Putting external heavy weights on the package could damage the product.

Product  Instructions.

1. Select a well convex area (muscle) next to the area to be operated but at least 20 cm far from it, possibly on the forearm, on the thigh or on a whip. Eventually move the plate in order that the longer side is perpendicular to the area of the operation (see the picture). Avoid that the plate is superimposed.

2.NB Current should flow in parallel to the body surface and never pass through the thorax. The patient must be placed on a dry and perfectly isolated surface. The patient must be kept isolated from conductive parts and the operating bed must be earthed. Use dry gauzes to avoid that different areas of the body come into contact with each other.

3. Do not place the plate on wounds and scars, near bone projections, very adipose tissues, close to metallic prosthesis or to ECG electrodes

or on areas, where liquids can flow.

4.Do not place the plate on areas, where it can be subjected to pressure or under patient’s weight.

5.Shave the selected area, clean carefully and skim in order to remove eventual remaining of cream or cosmetics. Dry the selected area.

6.NB Do not use inflammables during the patient’s preparation. Do not use alcoholic substances or benzoine tincture.

7.Eventual shave remaining or hair can cause burns.

Application

  1.  After controlling the packing for damages, check the plates and gel. Do not use the product  in the case of damages or visible defects. Instead, return it to FIAB.
  2. Only open package immediately before use.
  3. Take off the protection film and place the plate on the prepared area; place it starting from a     side and continue with same pressing until the opposite side.
  4. Make sure that the entire surface of the plate is in contact with the patient’s skin.
  5. Avoid excessive contact between fingers and adhesive. Do not touch the gel . NB There must not be air bubbles or parts of the adhesive support not perfectly close to the patient.
  6. Connect the plate to the high-frequency generator using the connectorcable .   For neutral plate without connection cable: lift the connector lever and plug in the tongue.Make sure to plug in the tongue until the end, then close the lever. Then connect the generator to the pencil and the eventual foot control.

Warning

        1.  This is a disposable product – do not re-use.

        2.  The product is supplied not sterile – do not sterilize.

        3. Do not re-position or cut the plate, do not add gel.

        4. Do not use the product in case:

        5. The packaging is not complete

        6.There are evident damages on the plate or on the connection cable, Gel is not homogeneous or is dry.

        7.Do not submit a patient with implanted pacemaker to electrosurgical current without first consulting a cardiologist.

        8. The devices are to be used by trained healthcare professionals in electrosurgical procedures.

        9. Do not use paediatric plates for high-power procedures (such as trans-urethral resection – TUR).


Product detail pictures:

China Cheap price Electrode Wire - 3M Pre-Wire Disposable Neutral Electrode  – Quanding detail pictures

China Cheap price Electrode Wire - 3M Pre-Wire Disposable Neutral Electrode  – Quanding detail pictures

China Cheap price Electrode Wire - 3M Pre-Wire Disposable Neutral Electrode  – Quanding detail pictures

China Cheap price Electrode Wire - 3M Pre-Wire Disposable Neutral Electrode  – Quanding detail pictures

China Cheap price Electrode Wire - 3M Pre-Wire Disposable Neutral Electrode  – Quanding detail pictures

China Cheap price Electrode Wire - 3M Pre-Wire Disposable Neutral Electrode  – Quanding detail pictures

China Cheap price Electrode Wire - 3M Pre-Wire Disposable Neutral Electrode  – Quanding detail pictures

China Cheap price Electrode Wire - 3M Pre-Wire Disposable Neutral Electrode  – Quanding detail pictures


Related Product Guide:

We normally believe that one's character decides products' quality, the details decides products' high-quality ,while using the REALISTIC,EFFICIENT AND INNOVATIVE staff spirit for China Cheap price Electrode Wire - 3M Pre-Wire Disposable Neutral Electrode – Quanding , The product will supply to all over the world, such as: Bolivia, El Salvador, Southampton, We are trying our best to make more customers happy and satisfied. we sincerely hope to establish a good long-term business relationship with your esteemed company thought this opportunity,based on equal, mutual beneficial and win win business from now till the future.
  • Superb technology, perfect after-sales service and efficient work efficiency, we think this is our best choice.
    5 Stars By Rebecca from Uruguay - 2017.08.16 13:39
    Cooperate with you every time is very successful, very happy. Hope that we can have more cooperation!
    5 Stars By Grace from London - 2018.06.05 13:10